FDA Adverse Event Malfunction Summary report: N

7700

MDR report key: 2772188 · Received October 2, 2012

Report

Report Number
9680959-2012-01227
Event Type
Malfunction
Date Received
October 2, 2012
Date of Event
September 25, 2012
Report Date
October 2, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EG
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND REPLACED 2 BLOWN FUSES. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM WOULD NOT PRODUCE X-RAYS. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7700 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7700

Patients

Seq Age Sex Outcome Treatment
1