ENDOPATH** XCEL* BLADELESS TROCAR
Report
- Report Number
- 3005075853-2012-04532
- Event Type
- Malfunction
- Date Received
- October 4, 2012
- Date of Event
- September 28, 2012
- Report Date
- October 1, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GCJ
- PMA / PMN Number
- K032676
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE (A) WAS RETURNED IN GOOD CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS FUNCTIONALLY TESTED TO DETECT ANY LEAKING ISSUES. UPON EVALUATION OF THE DEVICE, IT WAS FUNCTIONALLY LEAK TESTED AND PASSED. THE DEVICE WAS FULLY FUNCTIONAL ACCORDING TO THE MANUFACTURING REQUIREMENTS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE ANALYSIS RESULTS FOUND THAT THE DEVICE (B) WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS FUNCTIONALLY TESTED TO DETECT ANY LEAKING ISSUES. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK DURING INSERTION AND REMOVAL OF THE TEST PROBE THROUGH THE DEVICE. IT SHOULD BE NOTED THAT AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE REPORTED INSUFFLATIONS ISSUES. THE FINAL QUALITY RELEASE CRITERIA WERE MET BEFORE THIS BATCH WAS RELEASED FOR DISTRIBUTION. DEVICE B ADDITIONAL INFORMATION: BATCH # G9L220, EXPIRATION DATE: 08/2015, MANUFACTURING DATE: 09/2010.
(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION: WERE ANY NOISES HEARD SUCH AS WHISTLING OR HISSING?---YES. IF SO, DID THE NOISE PREVENT INSUFFLATION? ---YES. WAS THERE A DROP IN PRESSURE? ---NO. IF YES, DID THIS AFFECT THE VISIBILITY OF THE SURGEON? WHAT WAS THE GAS CONSUMPTION RATE OR VOLUME (LITER/MINUTE)? ---NO INFORMATION. WERE YOU ABLE TO IDENTIFY WHERE THE LEAK WAS COMING FROM? ---OUTER SEAL WAS A DEVICE INSERTED IN THE TROCAR DURING THE LEAKING? ---NO INFORMATION. IF SO, WHAT DEVICE? WAS ANY TORQUE BEING APPLIED TO THE TROCAR OR DEVICE? ---NO INFORMATION.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC PANCREAS REMOVAL PROCEDURE, AIR LEAKED FROM THE OUTER SEAL. ALTHOUGH THE OUTER SEAL WAS REPLACE WITH THE 2ND ONE, AIR KEPT LEAKING. THE SAME DEVICE WAS USED AS IS TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH** XCEL* BLADELESS TROCAR | LAPAROSCOPE, GENERAL AND PLASTIC SURGERY | GCJ | ETHICON ENDO-SURGERY, LLC. | NA | G4TX9A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |