FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2771825 · Received October 2, 2012

Report

Report Number
1720753-2012-07856
Event Type
Malfunction
Date Received
October 2, 2012
Date of Event
September 25, 2012
Report Date
October 2, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HARD DRIVE AND THE SOFTWARE UPGRADE WERE INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THERE WAS NO PATIENT INJURY OR DEATH REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1