FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2771816 · Received October 2, 2012

Report

Report Number
1720753-2012-07853
Event Type
Malfunction
Date Received
October 2, 2012
Date of Event
September 24, 2012
Report Date
October 2, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HIGH VOLTAGE CABLE WAS CLEANED AND REGREASED. THE LEMO PIN CONNECTOR WAS REPLACED AND THE FILAMENT WAS CALIBRATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS A FILAMENT REGULATOR ERROR DURING A CASE. THERE WAS NO REPORT OF PATIENT INJURY OR DEATH.