FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2771768 · Received October 2, 2012

Report

Report Number
1720753-2012-07829
Event Type
Malfunction
Date Received
October 2, 2012
Date of Event
September 12, 2012
Report Date
October 2, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE IDENTIFIED OR DUPLICATED. THE SYSTEM WAS OPERATING AS INTENDED. HOWEVER, THE SERVICE REPRESENTATIVE REPLACED THE HARD DRIVE AND RELOADED THE SYSTEM SOFTWARE AS A PRECAUTIONARY MEASURE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM LOCKED UP. THERE IS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1