OVER-THE-WIRE EMBOLECTOMY CATHETER
Report
- Report Number
- 1220948-2012-00013
- Event Type
- Other
- Date Received
- September 26, 2012
- Date of Event
- August 28, 2012
- Report Date
- September 2, 2012
- Manufacturer
- LEMAITRE VASCULAR INC
- Product Code
- DXE
- PMA / PMN Number
- K022145
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
WE HAVE RECEIVED THE DEVICE FOR EVALUATION A ND WERE ABLE TO CONFIRM THE FAILURE: THE TIP OF THE CATHETER WAS DETACHED. ADDITIONALLY, THE LUMEN OF THE CATHETER WAS STRETCH/ELONGATED. THE MOST PROBABLE ROOT CAUSE OF THE FAILURE IS FROM EXCESSIVE FORCE USED DURING THE PROCEDURE. THE LEMAITRE VASCULAR IFU WARNS TO EXERCISE CAUTION WHEN ENCOUNTERING EXTREMELY DISEASED VESSELS AND AVOID USING EXCESSIVE FORCE TO PUSH OR PULL CATHETER AGAINST RESISTANCE. THE PHYSICIAN DID NOT TAKE INTO CONSIDERATION THESE WARNING WHILE HE PERFORMED THE PROCEDURE. A LOT HISTORY INVESTIGATION WAS PERFORMED. THE DEVICE HISTORY RECORD REVIEW DID NOT REVEAL ANY DISCREPANCIES RELATED TO THE COMPLAINT EVENT DURING THE MANUFACTURING AND PACKAGING PROCESSES AND THERE WERE NO UNRESOLVED ISSUES RELATED TO THE COMPLAINT EVENT. IN ADDITION, PLEASE NOTE THAT THE PT IS OKAY.
THE PHYSICIAN PERFORMED THROMBECTOMY. THE PHYSICIAN FELT UNK RESISTANCE WHEN HE PULLED THE CATHETER WITH THE INFLATED BALLOON (PERFORMED THROMBECTOMY). THE TIP OF THE CATHETER (THE BALLOON AREA) WAS DETACHED WHEN THE SURGEON PULLED THE CATHETER OUT OF THE PT. THE TIP OF THE CATHETER WAS SUBSEQUENTLY REMOVED FROM THE PT ON (B)(6) 2012. THE PT IS RECOVERED WITHOUT FURTHER INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OVER-THE-WIRE EMBOLECTOMY CATHETER | DUAL LUMEN EMBOLECTOMY CATHETER | DXE | LEMAITRE VASCULAR INC | NA | OTW2017 OR OTW2035 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |