FDA Adverse Event Other Summary report: N

OVER-THE-WIRE EMBOLECTOMY CATHETER

MDR report key: 2771721 · Received September 26, 2012

Report

Report Number
1220948-2012-00013
Event Type
Other
Date Received
September 26, 2012
Date of Event
August 28, 2012
Report Date
September 2, 2012
Manufacturer
LEMAITRE VASCULAR INC
Product Code
DXE
PMA / PMN Number
K022145
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

WE HAVE RECEIVED THE DEVICE FOR EVALUATION A ND WERE ABLE TO CONFIRM THE FAILURE: THE TIP OF THE CATHETER WAS DETACHED. ADDITIONALLY, THE LUMEN OF THE CATHETER WAS STRETCH/ELONGATED. THE MOST PROBABLE ROOT CAUSE OF THE FAILURE IS FROM EXCESSIVE FORCE USED DURING THE PROCEDURE. THE LEMAITRE VASCULAR IFU WARNS TO EXERCISE CAUTION WHEN ENCOUNTERING EXTREMELY DISEASED VESSELS AND AVOID USING EXCESSIVE FORCE TO PUSH OR PULL CATHETER AGAINST RESISTANCE. THE PHYSICIAN DID NOT TAKE INTO CONSIDERATION THESE WARNING WHILE HE PERFORMED THE PROCEDURE. A LOT HISTORY INVESTIGATION WAS PERFORMED. THE DEVICE HISTORY RECORD REVIEW DID NOT REVEAL ANY DISCREPANCIES RELATED TO THE COMPLAINT EVENT DURING THE MANUFACTURING AND PACKAGING PROCESSES AND THERE WERE NO UNRESOLVED ISSUES RELATED TO THE COMPLAINT EVENT. IN ADDITION, PLEASE NOTE THAT THE PT IS OKAY.

Description of Event or Problem · 1

THE PHYSICIAN PERFORMED THROMBECTOMY. THE PHYSICIAN FELT UNK RESISTANCE WHEN HE PULLED THE CATHETER WITH THE INFLATED BALLOON (PERFORMED THROMBECTOMY). THE TIP OF THE CATHETER (THE BALLOON AREA) WAS DETACHED WHEN THE SURGEON PULLED THE CATHETER OUT OF THE PT. THE TIP OF THE CATHETER WAS SUBSEQUENTLY REMOVED FROM THE PT ON (B)(6) 2012. THE PT IS RECOVERED WITHOUT FURTHER INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OVER-THE-WIRE EMBOLECTOMY CATHETER DUAL LUMEN EMBOLECTOMY CATHETER DXE LEMAITRE VASCULAR INC NA OTW2017 OR OTW2035

Patients

Seq Age Sex Outcome Treatment
1 Other