ANIMAS VIBE
Report
- Report Number
- 2531779-2012-11796
- Event Type
- Malfunction
- Date Received
- October 4, 2012
- Report Date
- September 7, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW UP #1 SUBMITTED: (B)(4) 2012. DEVICE EVALUATION: THE INSULIN PUMP HAS BEEN RETURNED AND ADDITIONAL INVESTIGATION WAS PERFORMED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: ON INVESTIGATION, THE KEYPAD WAS FOUND TO BE FULLY INTACT WITHOUT PEELING OR VISIBLE DAMAGE. ON TESTING, ALL THE KEYPAD BUTTONS HAD INTERMITTENT RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS.
ON (B)(6) 2012, THE DISTRIBUTOR CONTACTED ANIMAS STATING THAT THE OK, UP ARROW AND DOWN ARROW KEYPAD BUTTONS WERE UNRESPONSIVE. THERE WAS NO ADDITIONAL INFORMATION AVAILABLE FOR THIS COMPLAINT. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANIMAS VIBE INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |