FDA Adverse Event Malfunction Summary report: N

HUDSON NEONATAL CIRCUIT, DUAL HTD LIMB, 60 IN

MDR report key: 2771652 · Received September 26, 2012

Report

Report Number
3004365956-2012-00293
Event Type
Malfunction
Date Received
September 26, 2012
Date of Event
September 10, 2012
Report Date
September 18, 2012
Manufacturer
TELEFLEX
Product Code
BZO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A DEVICE HISTORY RECORD (DHR) REVIEW COULD NOT BE PERFORMED AS THE LOT NUMBER WAS UNK. THE SAMPLE WAS RECEIVED FOR EVALUATION. A VISUAL EXAM WAS PERFORMED AND IT WAS FOUND THAT THE EXPIRATORY LIMB HAD A 1-INCH MELT, 25-INCHES FROM THE PT WYE. THE INSPIRATORY LIMB HAD NO NOTICEABLE DEFECTS. THE TUBING APPEARS TO HAVE BEEN FLATTENED AT THE LOCATION OF THE MELT. THIS INDICATES IT MAY HAVE BEEN PINCHED OR BECAME WEDGED BETWEEN TWO OBJECTS. THE RESISTANCE OF THE INSPIRATORY AND EXPIRATORY WIRES WERE MEASURED AND BOTH WERE FOUND TO BE WITHIN ALLOWABLE RANGE. BASED ON THE INVESTIGATION PERFORMED, THE COMPLAINT OF MELTED CIRCUIT WAS CONFIRMED, HOWEVER, THE ROOT CAUSE COULD NOT BE DETERMINED. POSSIBLE CAUSE OF A MELTED CIRCUIT INCLUDE NO FLOW, LOW FLOW, PROBE OUT, COVERED CIRCUITS, OR KINKED CIRCUITS. THE PRODUCT INSTRUCTIONS FOR USE(IFU) ALSO STATES SEVERAL WARNINGS IN ORDER TO AVOID OVERHEATING THE CIRCUIT. NO CONCLUSION CODE COULD BE CHOSEN AS THE COMPLAINT IS CONFIRMED, HOWEVER, A ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

COMPLAINT ALLEGES THAT THE CIRCUIT MELTED ON EXHALATION SIDE ONLY. PT AIRWAY TEMPERATURE SAID TO BE AT 37 DEGREES WITH NORMAL CONDENSATION. THE MELTED SPOT IS LOCATED IN ONLY ONE LOCATION ON EXPIRATORY CIRCUIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HUDSON NEONATAL CIRCUIT, DUAL HTD LIMB, 60 IN NEONATAL HEATED WIRE CIRCUIT BZO TELEFLEX

Patients

Seq Age Sex Outcome Treatment
1