HUDSON NEONATAL CIRCUIT, DUAL HTD LIMB, 60 IN
Report
- Report Number
- 3004365956-2012-00293
- Event Type
- Malfunction
- Date Received
- September 26, 2012
- Date of Event
- September 10, 2012
- Report Date
- September 18, 2012
- Manufacturer
- TELEFLEX
- Product Code
- BZO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
A DEVICE HISTORY RECORD (DHR) REVIEW COULD NOT BE PERFORMED AS THE LOT NUMBER WAS UNK. THE SAMPLE WAS RECEIVED FOR EVALUATION. A VISUAL EXAM WAS PERFORMED AND IT WAS FOUND THAT THE EXPIRATORY LIMB HAD A 1-INCH MELT, 25-INCHES FROM THE PT WYE. THE INSPIRATORY LIMB HAD NO NOTICEABLE DEFECTS. THE TUBING APPEARS TO HAVE BEEN FLATTENED AT THE LOCATION OF THE MELT. THIS INDICATES IT MAY HAVE BEEN PINCHED OR BECAME WEDGED BETWEEN TWO OBJECTS. THE RESISTANCE OF THE INSPIRATORY AND EXPIRATORY WIRES WERE MEASURED AND BOTH WERE FOUND TO BE WITHIN ALLOWABLE RANGE. BASED ON THE INVESTIGATION PERFORMED, THE COMPLAINT OF MELTED CIRCUIT WAS CONFIRMED, HOWEVER, THE ROOT CAUSE COULD NOT BE DETERMINED. POSSIBLE CAUSE OF A MELTED CIRCUIT INCLUDE NO FLOW, LOW FLOW, PROBE OUT, COVERED CIRCUITS, OR KINKED CIRCUITS. THE PRODUCT INSTRUCTIONS FOR USE(IFU) ALSO STATES SEVERAL WARNINGS IN ORDER TO AVOID OVERHEATING THE CIRCUIT. NO CONCLUSION CODE COULD BE CHOSEN AS THE COMPLAINT IS CONFIRMED, HOWEVER, A ROOT CAUSE COULD NOT BE DETERMINED.
COMPLAINT ALLEGES THAT THE CIRCUIT MELTED ON EXHALATION SIDE ONLY. PT AIRWAY TEMPERATURE SAID TO BE AT 37 DEGREES WITH NORMAL CONDENSATION. THE MELTED SPOT IS LOCATED IN ONLY ONE LOCATION ON EXPIRATORY CIRCUIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HUDSON NEONATAL CIRCUIT, DUAL HTD LIMB, 60 IN | NEONATAL HEATED WIRE CIRCUIT | BZO | TELEFLEX |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |