FDA Adverse Event
Malfunction
Summary report: N
TPS MICRO DRIVER
MDR report key: 2771446
·
Received September 28, 2012
Report
- Report Number
- 1811755-2012-03692
- Event Type
- Malfunction
- Date Received
- September 28, 2012
- Date of Event
- September 12, 2012
- Report Date
- September 12, 2012
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE MOTOR, THE PC BOARD AND OTHER PARTS WERE DISCOVERED TO BE THIRD PARTY PARTS. THIS UNAUTHORIZED MODIFICATION OF THESE COMPONENTS CAN LEAD TO THE REPORTED BIAS CURRENT ERROR. THE DEVICE HAS NOT BEEN RETURNED TO THE USER FACILITY AT THIS TIME.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE DEVICE CAUSED AS BIAS CURRENT ERROR TO BE DISPLAYED ON THE CONSOLE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TPS MICRO DRIVER | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |