FDA Adverse Event Malfunction Summary report: N

REMB SAG SAW

MDR report key: 2771444 · Received September 28, 2012

Report

Report Number
1811755-2012-03694
Event Type
Malfunction
Date Received
September 28, 2012
Date of Event
September 12, 2012
Report Date
September 12, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CORROSION WAS FOUND WITHIN THE MOTOR AND THE PRINTED CIRCUIT BOARD (FLEX ASSEMBLY) WITHIN THE MOTOR WAS FOUND TO BE DAMAGED, WHICH CAN CONTRIBUTE TO DEVICES OPERATING WITHOUT USER ACTIVATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT THE USER FACILITY THE SAW RAN WITHOUT USER ACTIVATION WHILE IT WAS IN SAFE MODE. THE USER FACILITY WAS NOT ABLE TO PROVIDE ANY FURTHER INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REMB SAG SAW HBE STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK