FDA Adverse Event Malfunction Summary report: N

MICRO SAGITTAL SAW

MDR report key: 2771441 · Received September 28, 2012

Report

Report Number
1811755-2012-03697
Event Type
Malfunction
Date Received
September 28, 2012
Date of Event
September 12, 2012
Report Date
September 12, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CORROSION WAS DISCOVERED IN THE MOTOR AND THE SOCKETS, WHICH IS A PROBABLE CAUSE OF THE DEVICE RUNNING WITHOUT USER ACTIVATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE SAW RAN WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO SAGITTAL SAW HBE STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK