FDA Adverse Event Malfunction Summary report: N

CORE SUMEX DRILL

MDR report key: 2771426 · Received September 28, 2012

Report

Report Number
1811755-2012-03706
Event Type
Malfunction
Date Received
September 28, 2012
Date of Event
September 3, 2012
Report Date
September 3, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

INTERNAL COMPONENTS WERE EVALUATED WHICH REVEALED THAT THE END OF MOTOR WAS WORN AND THE COIL WAS BROKEN.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING AT THE MFR, THE DEVICE DISPLAYED A BIAS CURRENT ERROR. THERE WAS O PT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE SUMEX DRILL ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK