FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2771420
·
Received September 28, 2012
Report
- Report Number
- 1811755-2012-03681
- Event Type
- Malfunction
- Date Received
- September 28, 2012
- Date of Event
- August 9, 2012
- Report Date
- September 11, 2012
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DUE TO THE DEVICE NOT POWERING UP DURING EVAL AT THE MFR, THE CUSTOMER'S COMPLAINT REGARDING OVERHEATING COULD NOT BE CONFIRMED. HOWEVER, DAMAGED BEARINGS AND CORROSION WITHIN THE MOTOR WERE DISCOVERED, WHICH CAN CAUSE OVERHEATING.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN EXTRACTION PROCEDURE AT THE USER FACILITY, THE DRILL WAS HEATING UP. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.