FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2771420 · Received September 28, 2012

Report

Report Number
1811755-2012-03681
Event Type
Malfunction
Date Received
September 28, 2012
Date of Event
August 9, 2012
Report Date
September 11, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DUE TO THE DEVICE NOT POWERING UP DURING EVAL AT THE MFR, THE CUSTOMER'S COMPLAINT REGARDING OVERHEATING COULD NOT BE CONFIRMED. HOWEVER, DAMAGED BEARINGS AND CORROSION WITHIN THE MOTOR WERE DISCOVERED, WHICH CAN CAUSE OVERHEATING.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN EXTRACTION PROCEDURE AT THE USER FACILITY, THE DRILL WAS HEATING UP. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK