Narratives

Additional Manufacturer Narrative · 1

FOLLOW UP # 1/SUPPLEMENTAL REPORT TEXT- 11/19/2012: THE LAY USER/PATIENT'S PRODUCT(S) HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER INVOLVED IN THIS CASE FAILED TESTING. THE METER WAS FOUND TO HAVE A DAMAGED DISPLAY. CRACKED / BROKEN LINES AND BLACK MARKS FOUND ON THE LCD. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE CRACKED DISPLAY WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT PING METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN INC. 3345659

Patients

Seq Age Sex Outcome Treatment
1 111 YR