FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2771135 · Received October 3, 2012

Report

Report Number
3004209178-2012-89851
Event Type
Malfunction
Date Received
October 3, 2012
Date of Event
October 1, 2012
Report Date
October 1, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS EXPERIENCING HIGH BLOOD GLUCOSE AND THE INSULIN PUMP DID NOT ALARM NO DELIVERY ALARM. IT WAS STATED THAT THE CUSTOMER GAVE A BOLUS OF 10.0 UNITS WHILE DISCONNECTED TO THE BODY, AND JUST TWO DROPS OF INSULIN EXITED. IT WAS STATED THAT THE INFUSION SET AND RESERVOIR WERE CHANGED SEVERAL TIMES, BUT ANOMALY PERSISTS. NO FURTHER INFORMATION WAS PROVIDED.