FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 2771103 · Received October 3, 2012

Report

Report Number
3004209178-2012-89871
Event Type
Malfunction
Date Received
October 3, 2012
Date of Event
March 2, 2012
Report Date
September 27, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
LZG
PMA / PMN Number
K053177
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. AN ERROR ALARM NOTED IN THE ALARM HISTORY AND THE DEVICE WAS RECEIVED WITH MOTOR ERROR ALARM STUCK IN THE BOLUS DELIVERY LOOP AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE. THE INSULIN PUMP WAS RECEIVED WITH SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED MOTOR ERROR DURING THE REWIND PROCESS. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM INSULIN INFUSION PUMP LZG LZG MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-712EWS

Patients

Seq Age Sex Outcome Treatment
1