OT ULTRALINK METER
Report
- Report Number
- 2939301-2012-11379
- Event Type
- Malfunction
- Date Received
- October 3, 2012
- Report Date
- September 20, 2012
- Manufacturer
- LIFESCAN INC.
- Product Code
- NBW
- PMA / PMN Number
- K073231
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
FOLLOW UP #2 (11/13/2012) DEVICE EVALUATION: THE SUBJECT TEST STRIPS FAILED TESTING, ON PERFORMANCE TESTING WITH LEVEL 2 CONTROL SOLUTION, THE TEST STRIPS WERE FOUND TO BE READING HIGH.
FOLLOW UP #1 (11/13/2012) DEVICE EVALUATION: THE METER AND TEST STRIPS INVOLVED IN THIS COMPLAINT HAVE BEEN RETURNED TO LFS AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 AND (B)(4) 2012 RESPECTIVELY WITH THE FOLLOWING FINDINGS: THE METER HAS PASSED ALL TESTING WITH NO FAULTS FOUND, THE COMPLAINT WAS NOT CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A SECOND FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE A PATIENT CONTACTED LIFESCAN ALLEGING THAT THE SUBJECT METER READ INACCURATELY HIGH COMPARED TO ANOTHER METER. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "196 MG/DL" WITH A LIFESCAN METER AND "132 MG/DL" ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <= 30% AND/OR <= 30 MG/DL. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.