LIGACLIP** ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 3005075853-2012-04519
- Event Type
- Malfunction
- Date Received
- October 3, 2012
- Date of Event
- September 20, 2012
- Report Date
- September 27, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION: WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON? ASKU. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? BILE DUCT. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? YES. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? ASKU. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? NO. WERE ANY UNEXPECTED NOISES HEARD? ASKU IF SO, WHEN? NO. DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? NO. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? NO. WHAT WERE THE INDICATIONS FOR SURGERY? WHAT WAS FOUND? CHOLECYSTITIS. DOES THE PATIENT HAVE ANY RELEVANT HISTORY OF SURGICAL TREATMENTS? IF SO, WHAT? UNKNOWN. DID SOMEBODY OTHER THAN THE PRIMARY SURGEON FIRE THE INSTRUMENT? IF SO, WHOM? NO.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED AND FED AND FORMED FOUR SCISSORED CLIPS, AND THEN THE REMAINING CLIPS WERE FORMED AS INTENDED. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED. ALTHOUGH IT IS NOT POSSIBLE TO CONCLUDE HOW THE CIRCUMSTANCES OCCURRED, IT IS KNOWN FROM THE HISTORY OF THE INSTRUMENT THAT AN INCORRECT/EXCESSIVE APPLICATION OF TORQUE TO THE JAWS DURING INSTRUMENT USE CREATES A MISALIGNMENT OF THE TIPS. IN ADDITION AS PER THE INSTRUCTIONS FOR USE "DO NOT EXCESSIVELY TWIST OR TORQUE THE INSTRUMENT JAWS WHEN POSITIONING THE INSTRUMENT ON A VESSEL AND FIRING. EXCESSIVE TWISTING OR TORQUING MAY RESULT IN CLIP MALFORMATION". THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THAT THE CLIPS WERE SCISSORING WHEN FIRED ON THE BILE DUCT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT ANY ISSUES. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | J4C510 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |