FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2770876 · Received October 3, 2012

Report

Report Number
1416980-2012-01483
Event Type
Malfunction
Date Received
October 3, 2012
Date of Event
September 11, 2012
Report Date
September 12, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED PROBLEM OF " BROKEN CABLE" WAS CONFIRMED DURING DEVICE EVALUATION. A NEW AC POWER CORD WAS INSTALLED TO FIX THE REPORTED PROBLEM

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP WITH A BROKEN CABLE. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1