FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 2770702 · Received October 1, 2012

Report

Report Number
9680959-2012-01214
Event Type
Malfunction
Date Received
October 1, 2012
Date of Event
September 24, 2012
Report Date
October 1, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HAND SWITCH WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM LOCKED UP AND CONTINUED TO X-RAY PRIOR TO A CASE. THE SYSTEM HAD TO BE REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER SYSTEM. THERE WAS NO PT INJURY OR DEATH

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1