FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2770642 · Received October 1, 2012

Report

Report Number
1720753-2012-07784
Event Type
Malfunction
Date Received
October 1, 2012
Date of Event
September 5, 2012
Report Date
October 1, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE FILAMENT DRIVER BOARD WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM KV RAMPED TO MAX AND AS A RESULT, THE LIVE IMAGE WAS UNUSABLE. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1