FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX* TISSUE MORCELLATOR

MDR report key: 2770549 · Received October 3, 2012

Report

Report Number
2210968-2012-05968
Event Type
Malfunction
Date Received
October 3, 2012
Report Date
September 12, 2012
Manufacturer
ETHICON, INC.
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED HAS BEEN RECEIVED, HOWEVER, THE PRODUCT EVALUATION IS NOT YET COMPLETE. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA. THIS IS ONE OF TWO MEDWATCHES BEING SUBMITTED AS TWO DEVICES WERE INVOLVED IN THIS EVENT. SEE ALSO MEDWATCH 2210968-2012-05967. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.

Additional Manufacturer Narrative · 1

DATE SENT TO THE FDA: (B)(4) 2012. IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY. DURING THE PROCEDURE, THE LIGHTS ON THE GENERATOR BEGAN TO FLASH AND THE DEVICE WOULD NOT CUT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES. THIS IS ONE OF THREE MEDWATCHES BEING SUBMITTED AS THREE DEVICES WERE INVOLVED IN THIS EVENT. SEE ALSO MEDWATCH 2210968-2012-05967 AND MEDWATCH 2210968-2012-08319. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.

Additional Manufacturer Narrative · 1

(B)(4)

Additional Manufacturer Narrative · 1

(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE BLADE FAILED TO ROTATE DURING THE EVALUATION. IT IS LIKELY THAT THE ACCUMULATION OF BLOOD, BODY FLUIDS, AND TISSUE PREVENTED THE DEVICE FROM CONTINUING TO OPERATE AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY. DURING THE PROCEDURE, THE LIGHTS ON THE GENERATOR BEGAN TO FLASH AND THE DEVICE STOPPED CUTTING INTERMITTENTLY. THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GYNECARE MORCELLEX* TISSUE MORCELLATOR LAPAROSCOPE, GYNECOLOGIC HET ETHICON, INC. NA MT216697