LEAD MODEL 302
Report
- Report Number
- 1644487-2012-02501
- Event Type
- Malfunction
- Date Received
- October 3, 2012
- Date of Event
- September 4, 2012
- Report Date
- September 10, 2012
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
MANUFACTURER REVIEW OF X-RAYS OF IMPLANTED DEVICE. X-RAYS REVIEWED BY THE MANUFACTURER. NO GROSS LEAD DISCONTINUITIES VISUALIZED. DEVICE FAILURE IS SUSPECTED BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
DEVICE MALFUNCTION SUSPECTED BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH.
THE PATIENT HAD REVISION SURGERY (B)(6) 2012. THEIR EXPLANTED LEAD HAS BEEN RETURNED FOR ANALYSIS AND COMPLETION IS PENDING. AN ENGINEER WAS PRESENT AT THE OPERATING ROOM ON (B)(6) 2012. THE LEAD IMPEDANCE WAS ASSESSED THROUGH SYSTEM DIAGNOSTICS AND IT WAS FOUND TO BE HIGH, WITH DCDC CODE 7 AND OUTPUT STATUS LIMIT. THE PULSE GENERATOR WAS EXPLANTED AND A REPLACEMENT GENERATOR WAS CONNECTED TO THE ORIGINALLY IMPLANTED LEAD. THE LEAD IMPEDANCE WAS ASSESSED AGAIN, THROUGH SYSTEM DIAGNOSTICS OF THE REPLACEMENT GENERATOR. THE AFOREMENTIONED RESULTS (LEAD IMPEDANCE HIGH, OUTPUT STATUS LIMIT, DCDC CODE 7) WERE OBTAINED AGAIN AND A LEAD REPLACEMENT WAS INITIATED. UPON CONNECTION OF THE REPLACEMENT GENERATOR TO THE REPLACEMENT LEAD, SYSTEM DIAGNOSTICS WERE RUN AND THE RESULTS WERE WITHIN NORMAL RAGES (DDC CODE 1, LEAD IMPEDANCE OK, OUTPUT STATUS OK). THE EXPLANTED GENERATOR WAS KEPT BY THE IMPLANTING HOSPITAL AND IT WILL BE DISPOSED OF IN THEIR OWN FACILITIES. FIBROSIS WAS NOTED AT THE ELECTRODE SITES. NO CAUSE OTHER THAN THE DEVICE IS KNOWN AT THIS TIME.
ADDITIONAL INFORMATION WAS RECEIVED ON (B)(6) 2012 WHEN PRODUCT ANALYSIS WAS COMPLETED ON THE EXPLANTED LEADS. THE ELECTRODES WERE NOT RETURNED FOR ANALYSIS; THEREFORE A COMPLETE EVALUATION COULD NOT BE PERFORMED ON THE ENTIRE LEAD PRODUCT. DURING THE VISUAL ANALYSIS THE CONNECTOR PIN QUADFILAR COIL APPEARED TO BE BROKEN APPROXIMATELY 337 MM FROM THE END OF THE CONNECTOR BOOT. SCANNING ELECTRON MICROSCOPY WAS PERFORMED ON AND IDENTIFIED THE AREA AS HAVING EXTENSIVE PITTING AND MECHANICAL DAMAGE WHICH PREVENTED IDENTIFICATION OF THE COIL FRACTURE TYPE. IT IS BELIEVED THAT STIMULATION WAS PRESENT FOR A CERTAIN PERIOD OF TIME AS EVIDENCED BY THE PRESENCE OF METAL PITTING. LOW MAGNIFICATION SEM ANALYSIS OF THE QUADFILAR COIL SHOWS CHARACTERISTICS TYPICAL OF A LEAD DISCONTINUITY WHICH MAY INCLUDE: MATERIAL FRACTURE, ROUGH OR PITTED SURFACE, THINNED MATERIAL THICKNESS, ELECTRO-ETCHING OR MATERIAL DISSOLUTION. WHAT APPEARED TO BE INCLUSIONS WERE OBSERVED ON SOME OF THE COIL STRANDS. EDS (ENERGY DISPERSION SPECTROSCOPY - PROVIDES CHEMICAL OR ELEMENT IDENTITY/COMPOSITION ANALYSIS) WAS PERFORMED ON THE INCLUSIONS. THE ANALYSIS RESULTS SHOW THE PRESENCE OF ELEMENTS CONSISTENT WITH THE MP35N MATERIAL. DURING THIS ANALYSIS THE EXACT IMPACT THE INCLUSIONS MADE TO THE OBSERVED FRACTURE COULD NOT BE DETERMINED. HOWEVER, RESULTS OF AN ENERGY DISPERSION SPECTROSCOPY (EDS) EVALUATION DEMONSTRATE THAT THE COMPOSITION OF MATERIALS IN THESE AREAS ARE REPRESENTATIVE OF THE COIL MATERIAL, SO THE CONDITION IS NOT EXPECTED TO HAVE AN ADVERSE EFFECT. THE ABRADED OPENING FOUND ON THE OUTER SILICONE TUBING, MOST LIKELY PROVIDED THE LEAKAGE PATH FOR WHAT APPEARED TO BE REMNANTS OF DRIED BODY FLUIDS FOUND INSIDE THE OUTER SILICONE TUBING. WITH THE EXCEPTION OF THE OBSERVED DISCONTINUITY, THE CONDITION OF THE RETURNED LEAD PORTION IS CONSISTENT WITH CONDITIONS THAT TYPICALLY EXIST FOLLOWING AN EXPLANT PROCEDURE. NO OTHER OBVIOUS ANOMALIES WERE NOTED. THE SETSCREW MARKS FOUND ON THE MARKED CONNECTOR PIN PROVIDE EVIDENCE THAT, AT ONE POINT IN TIME, A GOOD MECHANICAL AND ELECTRICAL CONNECTION WAS PRESENT. CONTINUITY CHECKS OF THE RETURNED LEAD PORTION WERE PERFORMED, DURING THE VISUAL ANALYSIS, AND NO OTHER DISCONTINUITIES WERE IDENTIFIED.
IT WAS REPORTED TO OUR COUNTRY REPRESENTATIVE IN (B)(6) THAT THERE WAS A PATIENT WITH HIGH LEAD IMPEDANCE. THE PATIENT'S GENERATOR WAS INTERROGATED ON (B)(4) 2012. SYSTEM DIAGNOSTIC TESTING RESULTED IN A DCDC=2 AND DIAGNOSTICS NORMAL MODE DCDC=4.. HIGH LEAD IMPEDANCE, DCDC 7 WAS DETECTED ON (B)(6) 201,2 DURING THE DEVICE'S INTERROGATION. THEIR DEVICE SWITCHED OFF AT THIS TIME. NO FURTHER INFORMATION HAS BEEN ATTAINED AT THIS TIME.
ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT'S HIGH IMPEDANCE WAS CAUSED BY POSSIBLE TRAUMA, SINCE THE PATIENT EXPERIENCES FREQUENT DROPS. THE PATIENT WILL BE SEEN BY A SURGEON IN THE COMING WEEKS. NO DATE SET AT THIS TIME. AP AND LATERAL CHEST AND AP AND LATERAL NECK X-RAYS WERE RECEIVED FOR REVIEW. REVIEW OF THE X-RAYS INDICATED THE FOLLOWING: THE GENERATOR IS PLACED IN A NORMAL LOCATION IN THE UPPER LEFT CHEST. THE FILTER FEED-THROUGH WIRES APPEARED TO BE INTACT. THE LEAD CONNECTOR PIN APPEARED TO BE FULLY INSERTED INTO THE GENERATOR CONNECTOR BLOCK. THE LEAD WIRES AT THE CONNECTOR PIN APPEARED TO BE INTACT. THE NEGATIVE AND POSITIVE ELECTRODE APPEARED TO BE CORRECTLY ALIGNED ON THE VAGAL NERVE. A STRAIN-RELIEF BEND WAS PRESENT. NO STRAIN-RELIEF LOOP WAS PRESENT. ONE TIE-DOWN HAD BEEN USE, PLACED A FEW CENTIMETERS FURTHER TOWARDS THE GENERATOR, TO HOLD THE UPPER THIRD OF THE LEAD BODY. NO TIE-DOWN WAS USED TO HOLD THE STRAIN-RELIEF BEND. PER LABELING, A STRAIN RELIEF BEND SHOULD BE PLACED STARTING 3 CM FROM THE ANCHOR TETHER AND SHOULD BE SECURED WITH A TIE-DOWN PARALLEL TO THE ANCHOR TETHER. A LARGE LOOP SHOULD THEN BE FORMED AND SECURED WITH AN ADDITIONAL TIE-DOWN TO ALLOW FOR FULL NECK MOVEMENT IN ALL DIRECTIONS. PORTIONS OF THE LEAD APPEARED TO BE BEHIND THE GENERATOR AND COULD NOT BE FULLY ASSESSED. NO GROSS FRACTURES OR SHARP ANGLES COULD BE IDENTIFIED. BASED ON THE X-RAYS IMAGES, NO EVIDENT CAUSE FOR THE HIGH IMPEDANCE COULD BE IDENTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LEAD MODEL 302 | LEAD MODEL 302 | LYJ | CYBERONICS, INC. | 302 | 2423 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |