CHARGING SYSTEM
Report
- Report Number
- 1627487-2012-01774
- Event Type
- Injury
- Date Received
- October 1, 2012
- Date of Event
- September 8, 2012
- Report Date
- September 8, 2012
- Manufacturer
- ST JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS - POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGES WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REF MFR REPORT: 1627487-2012-1773. IT WAS REPORTED, THE PT'S IPG TURNED OFF BY ITSELF AND WAS UNABLE TO TURN ON WITH THE MAGNET. IT WAS ALSO REPORTED, THE PT IS EXPERIENCING POCKET HEATING AT THE IPG SITE WHILE CHARGING. THE PT WAS ADVISED TO CHARGE FOR SHORTER DURATIONS AND MORE FREQUENTLY. IT WAS ALSO REPORTED, THE PT IS EXPERIENCING PAIN AT THE IPG SITE. X-RAYS HAVE BEEN ORDERED AND F/U IS PENDING WITH HIS PHYSICIAN. SURGICAL INTERVENTION IS PENDING TO ADDRESS THE ISSUE. ON 08/01/2012 ST JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CHARGING SYSTEM | SCS CHARGER | LGW | ST JUDE MEDICAL - NEUROMODULATION | 3721 | 3682650 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Required Intervention | SCS LEAD: MODEL 3186 (2)| IMPLANT DATE: |