FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 2770407 · Received October 1, 2012

Report

Report Number
1627487-2012-01774
Event Type
Injury
Date Received
October 1, 2012
Date of Event
September 8, 2012
Report Date
September 8, 2012
Manufacturer
ST JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS - POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGES WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REF MFR REPORT: 1627487-2012-1773. IT WAS REPORTED, THE PT'S IPG TURNED OFF BY ITSELF AND WAS UNABLE TO TURN ON WITH THE MAGNET. IT WAS ALSO REPORTED, THE PT IS EXPERIENCING POCKET HEATING AT THE IPG SITE WHILE CHARGING. THE PT WAS ADVISED TO CHARGE FOR SHORTER DURATIONS AND MORE FREQUENTLY. IT WAS ALSO REPORTED, THE PT IS EXPERIENCING PAIN AT THE IPG SITE. X-RAYS HAVE BEEN ORDERED AND F/U IS PENDING WITH HIS PHYSICIAN. SURGICAL INTERVENTION IS PENDING TO ADDRESS THE ISSUE. ON 08/01/2012 ST JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CHARGING SYSTEM SCS CHARGER LGW ST JUDE MEDICAL - NEUROMODULATION 3721 3682650

Patients

Seq Age Sex Outcome Treatment
1 51 YR Required Intervention SCS LEAD: MODEL 3186 (2)| IMPLANT DATE: