OT VERIO IQ METER
Report
- Report Number
- 3008382007-2012-04708
- Event Type
- Malfunction
- Date Received
- October 3, 2012
- Report Date
- September 7, 2012
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
(B)(4): THE METER HAS PASSED TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2012 THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) FROM (B)(6) ALLEGING THAT THEIR ONETOUCH VERIO IQ METER WAS PROMPTING A BATTERY INDICATOR SYMBOL INSTEAD OF THE APPLY SAMPLE SYMBOL WHEN THEY ATTEMPTED TO TEST THEIR BLOOD GLUCOSE. PER THE ONETOUCH VERIO IQ PRODUCT MANUAL THE BATTERY INDICATOR PROMPTS WHEN THE BATTERY NEEDS TO BE REPLACED IN THE METER. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY DUE TO THE ALLEGED ISSUE. THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. REPLACEMENT PRODUCT WAS SENT TO THE PATIENT. BASED ON THE INFORMATION PROVIDED, THEIR COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION(S): THE PATIENT CLAIMED THAT THE SUBJECT METER WAS PROMPTING A BATTERY INDICATOR INSTEAD OF THE APPLY SAMPLE SYMBOL WHEN THEY ATTEMPTED TO TEST THEIR BLOOD GLUCOSE. THERE IS NO EVIDENCE THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY SYMPTOMS, TREATMENT, OR BLOOD GLUCOSE VALUES SUGGESTIVE OF A SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3254033 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |