FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2770101 · Received October 3, 2012

Report

Report Number
2531779-2012-11688
Event Type
Malfunction
Date Received
October 3, 2012
Report Date
September 6, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 11/05/2012- DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE AUDIO BOLUS BUTTON WAS RETURNED TORN. A DAMAGED BUTTON WILL PERMIT CONTAMINATION TO PERMEATE WHICH WILL HAVE A NEGATIVE IMPACT ON BUTTON FUNCTION. THIS SITUATION IS NOT LIKELY TO RESULT IN AN ADVERSE EVENT AS THE DAMAGED BOLUS BUTTON SHOULD BE CLEARLY VISIBLE AND PROHIBIT THE USE. DURING TESTING, THE AUDIO BOLUS BUTTON RESPONDED APPROPRIATELY. UNRELATED TO THE COMPLAINT, THE KEYPAD WAS REMOVED AND EVIDENCE OF A MILD WHITE RESIDUE WAS PRESENT UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

THE DISTRIBUTOR CONTACTED ANIMAS ON (B)(6) 2012 STATING THAT THE KEYPAD WAS DAMAGED AND THE AUDIO BOLUS BUTTON WAS UNRESPONSIVE. THE DISTRIBUTOR WAS UNABLE TO DETERMINE WHETHER THE DAMAGE OR THE RESPONSIVENESS ISSUES HAD OCCURRED FIRST. THERE IS NO REPORTED ADVERSE EVENT WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD RESPONSIVENESS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1