FDA Adverse Event Injury Summary report: N

MBT CEM KEEL TIB TRAY SZ3

MDR report key: 2770098 · Received October 3, 2012

Report

Report Number
1818910-2012-20235
Event Type
Injury
Date Received
October 3, 2012
Date of Event
September 26, 2012
Report Date
September 26, 2012
Manufacturer
DEPUY IRELAND
Product Code
NJL
PMA / PMN Number
P830055/S108
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION SINCE ITS RELEASE FOR DISTRIBUTION. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A; REV. C. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED.DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS LOOSENING OF THE TIBIAL TRAY AT THE CEMENT/IMPLANT INTERFACE AND PAIN. DEPUY CEMENT WAS NOT USED AT THE TIME OF ORIGINAL IMPLANTATION. (RIGHT KNEE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MBT CEM KEEL TIB TRAY SZ3 TIBIAL KNEE PROSTHESIS NJL DEPUY IRELAND 3374666

Patients

Seq Age Sex Outcome Treatment
1 52 YR Required Intervention