FDA Adverse Event Malfunction Summary report: N

NEOPUFF INFANT RESUSCITATOR

MDR report key: 2769658 · Received October 2, 2012

Report

Report Number
9611451-2012-00703
Event Type
Malfunction
Date Received
October 2, 2012
Date of Event
September 13, 2012
Report Date
September 20, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTL
PMA / PMN Number
K892885
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT RD900AEU NEOPUFF INFANT RESUSCITATOR WAS RECEIVED FOR SERVICE AT OUR FISHER & PAYKEL HEALTHCARE (FPH) SERVICE CENTRE IN (B)(4) WHERE IT WAS INSPECTED BY A TRAINED FPH TECHNICIAN. THE RETURNED NEOPUFF UNIT WAS VISUALLY INSPECTED FOR DAMAGE AND PERFORMANCE TESTED IN ACCORDANCE WITH THE NEOPUFF TECHNICAL MANUAL. OUR ANALYSIS IS BASED ON THE SERVICE FINDINGS PROVIDED BY OUR SERVICE CENTRE. RESULTS: VISUAL INSPECTION REVEALED THAT THE PLUG SET WAS MISSING. NO OTHER PHYSICAL DAMAGE WAS NOTED TO THE RETURNED NEOPUFF UNIT. THE REPORTED FAULT WITH THE MANOMETER WAS NOT REPLICATED DURING PERFORMANCE CHECK. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 061012. CONCLUSION: NO FAULT WAS FOUND WITH THE MANOMETER OF THE RETURNED NEOPUFF UNIT. IF FUNCTIONED AS INTENDED DURING PERFORMANCE CHECK. THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANT UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF/PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT." FOLLOWING INSPECTION OF THE MANOMETER AND REPLACEMENT OF THE PLUG SET, THE NEOPUFF UNIT WAS RETURNED TO THE HOSPITAL.

Description of Event or Problem · 1

A HOSPITAL IN (B)(4) REPORTED THAT AN RD900 NEOPUFF INFANT RESUSCITATOR WOULD NOT CALIBRATE OR VERIFY THE MANOMETER PRESSURES. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEOPUFF INFANT RESUSCITATOR BTL BTL FISHER & PAYKEL HEALTHCARE LIMITED RD900AEU

Patients

Seq Age Sex Outcome Treatment
1