FDA Adverse Event Malfunction Summary report: N

LASSO® 2515 NAV ECO VARIABLE CATHETER

MDR report key: 2769184 · Received October 2, 2012

Report

Report Number
9673241-2012-00272
Event Type
Malfunction
Date Received
October 2, 2012
Date of Event
September 12, 2012
Report Date
September 12, 2012
Manufacturer
BIOSENSE WEBSTER, INC. (JUAREZ)
Product Code
DRF
PMA / PMN Number
K113213
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON FOLLOW-UP ATTEMPTS TO REQUEST FOR THE COMPLAINT PRODUCT TO BE RETURNED FOR ANALYSIS, ON (B)(4) 2012 WE WERE NOTIFIED THAT THE CUSTOMER HAD DISCARDED THE CATHETER INVOLVED IN THIS COMPLAINT. THE DEVICE HISTORY RECORD WAS REVIEWED, AND WAS VERIFIED THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. (B)(4).

Additional Manufacturer Narrative · 1

CONCOMITANT BWI PRODUCTS USED DURING THE PROCEDURE: CARTO 3 SYSTEM, MODEL #: M-4800-01 , SERIAL #: (B)(4). COOL FLOW IRRIGATION PUMP, MODEL #: M-5491-02, SERIAL #: (B)(4). STOCKERT RF GENERATOR, MODEL #:M-5463-01, SERIAL #: (B)(4). EZ STEER THERMOCOOL SF NAV CATHETER, MODEL #: D-1317-05-S, LOT #.: UNKNOWN. SOUNDSTAR 3D 10F-90 CATHETER, MODEL #: M-5723-05, LOT #.: UNKNOWN. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A FIB PROCEDURE A CLOT WAS FOUND ON THE LASSO NAV VARIABLE CATHETER. THE CLOT WAS ASSESSED WITH THE INTRACARDIAC ULTRASOUND. THE PHYSICIAN ATTEMPTED TO ASPIRATE THE CLOT, HOWEVER, THE RESULT WAS UNCERTAIN. THERE WAS NO DEFINITE CLOT IN THE ASPIRATE. LATER ON THE ICE DID NOT SHOW ANY MORE CLOT. THE CASE WAS CONTINUED AND AT THE TIME OF THE CALL, THE PATIENT WAS STABLE WITH NO INJURY. THE ACT RANGE DURING THE PROCEDURE WAS REPORTED TO BE >350.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LASSO® 2515 NAV ECO VARIABLE CATHETER CATHETER, ELECTRODE RECORDING OR PROBE, ELECTRODE RECORDING DRF BIOSENSE WEBSTER, INC. (JUAREZ) D-1343-02-S 15619962L

Patients

Seq Age Sex Outcome Treatment
1