FDA Adverse Event Malfunction Summary report: N

BABY CONTROL WALL MOUNT INFANT WARMER

MDR report key: 2766240 · Received October 1, 2012

Report

Report Number
9611451-2012-00687
Event Type
Malfunction
Date Received
October 1, 2012
Report Date
September 3, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
FMT
PMA / PMN Number
K971461
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4), SERIAL NUMBERS AND LOT DATE: (B)(4) 070130, (B)(4) - 070130, (B)(4) - 070130, (B)(4) - 070130, (B)(4) - 070130, (B)(4) - 070130. DEVICE MANUFACTURER DATE: 01/30/2007. METHOD: THE COMPLAINT WARMERS ARE NOT EXPECTED TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. OUR INVESTIGATION IS ACCORDINGLY BASED ON THE EVENT DESCRIPTION, PHOTOGRAPHS PROVIDED AND OUR KNOWLEDGE OF THE PRODUCT. RESULTS: INSPECTION OF THE PHOTOGRAPHS PROVIDED BY THE CUSTOMER REVEALED CRACKS ON THE DOME PORTION OF THE HEATER HEAD, CONFIRMING THE REPORTED FAULT. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR THIS LOT NUMBER. CONCLUSION: IT IS LIKELY THAT THE FLAKING AND CRACKING OF THE COMPLAINT WARMER HEAD IS RELATED TO A KNOWN PROBLEM OF INCOMPATIBLE CLEANING SOLUTIONS REACTING WITH THE POLYCARBONATE PLASTIC AND ACRYLIC COMPONENTS OF THE INFANT WARMER. WHEN THE HEATER HEAD BEGINS TO WARM UP DURING USE, A CHEMICAL REACTION WITH THE INCOMPATIBLE CLEANING SOLUTION RESULTS IN GRADUAL CRACKING AND CRAZING OF THE HEATER HEAD THE INFANT WARMER SERVICE MANUAL FEATURES A DIAGRAM OF THE INFANT WARMER WHICH HIGHLIGHTS POLYCARBONATE AND ACRYLIC COMPONENTS AND STATES THE FOLLOWING: "CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYPOCHLORITES, PEROXIDES, GLUTERALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE." "CAUTION THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES". AS PART OF CONTINUOUS IMPROVEMENT, WE HAVE SINCE REVISED OUR CLEANING INSTRUCTIONS TO RECOMMEND SPECIFIC PROPRIETARY CLEANING WIPES. A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE HAS SINCE BEEN IN CONTACT WITH THE CUSTOMER AND PROVIDED GUIDELINES ON CLEANING THE PRODUCTS AS PER THE TECHNICAL MANUAL.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THE HEATER HEAD OF SIX BABY CONTROL WALL MOUNT INFANT WARMERS WERE CRACKED. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BABY CONTROL WALL MOUNT INFANT WARMER FMT FMT FISHER & PAYKEL HEALTHCARE LTD IW980 SEE H.10

Patients

Seq Age Sex Outcome Treatment
1