INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Report
- Report Number
- 9611451-2012-00692
- Event Type
- Malfunction
- Date Received
- October 1, 2012
- Date of Event
- September 17, 2012
- Report Date
- September 17, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT268 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. METHOD: THE COMPLAINT BREATHING CIRCUIT WAS RETURNED TO THE MANUFACTURER. THE COMPLAINT BREATHING CIRCUIT WAS VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAKS. RESULTS: NO DAMAGES WERE OBSERVED ON THE RETURNED COMPLAINT BREATHING CIRCUIT. THE PRESSURE TEST IDENTIFIED THE COMPLAINT BREATHING CIRCUIT TO BE WITHIN SPECIFICATION FOR THIS PRODUCT. NO FAULT FOUND. CONCLUSION: WE WERE UNABLE TO DETERMINE THE CAUSE OF THE PROBLEM AS REPORTED BY THE CUSTOMER, AS NO FAULT COULD BE FOUND WITH THE RETURNED COMPLAINT DEVICE. ALL RT268 BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS PRIOR TO DISTRIBUTION AND THOSE THAT FAIL ARE REJECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT268 STATE THE FOLLOWING: - CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE. - PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. - SET APPROPRIATE VENTILATOR ALARMS.
A HOSPITAL IN (B)(6) REPORTED THAT AN RT268 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT DID NOT PASS THE VENTILATOR LEAK TEST. THE HOSPITAL FURTHER REPORTED THAT A NEW CIRCUIT WAS USED AND THE PROBLEM WAS RESOLVED. THIS WAS REPORTED PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT268 | 120622 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |