FDA Adverse Event Injury Summary report: N

HAKIM PROGRAMMABLE VALVE

MDR report key: 2763131 · Received September 27, 2012

Report

Report Number
1226348-2012-00481
Event Type
Injury
Date Received
September 27, 2012
Date of Event
August 30, 2012
Manufacturer
CODMAN AND SHURTLEFF - MEDOS
Product Code
JXG
PMA / PMN Number
K980778
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION OF THE VALVE REVEALED THAT THE PRODUCT CODE RETURNED WAS ACTUALLY THAT OF 82-3113 AND NOT 82-3113 AS INITIALLY REPORTED. IT WAS ALSO NOTED THAT THERE WAS THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICE. THIS BIOLOGICAL DEBRIS CAUSED AN OCCLUSION IN THE DEVICE AND FOR THE RETURNED VALVE TO FAIL THE PROGRAMMING TEST, WHICH APPEARS TO HAVE CAUSED THE DIFFICULTY ENCOUNTERED BY THE CUSTOMER. REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE VALVE MET SPECIFICATION REQUIREMENTS WHEN RELEASED TO STOCK. BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CLOSED.

Additional Manufacturer Narrative · 1

HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICES WHICH HAVE INTERFERED WITH THEIR PROGRAMMING AND/OR PRESSURE CONTROL MECHANISMS. THE BIOLOGICAL DEBRIS HAS CAUSED THE RETURNED VALVES TO FAIL THE PROGRAMMING AND/OR PRESSURE TESTS AND APPEARS TO HAVE CAUSED THE DIFFICULTIES EXPERIENCED BY THE CUSTOMERS. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFORMATION REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CONSIDERED TO BE CLOSED AT THIS TIME.

Description of Event or Problem · 1

AFFILIATE REPORTED THAT BEFORE THE SHUNT REVISION, THE VALVE COULD NOT BE REPROGRAMMED. THERE WERE 15 ATTEMPTS TO REPROGRAM USING X-RAY AND A PROGRAMMER. AS A RESULT THE DEVICE WAS REVISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HAKIM PROGRAMMABLE VALVE SHUNT, CENTRAL NERVOUS SYSTEM COMPS JXG CODMAN AND SHURTLEFF - MEDOS

Patients

Seq Age Sex Outcome Treatment
1 35 YR Required Intervention