ANIMAS INSULIN CARTRIDGE

MDR report key: 2761342 · Received September 26, 2012

Report

Report Number
2531779-2012-11325
Event Type
Injury
Date Received
September 26, 2012
Report Date
August 29, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP ((B)(4) 2012)-DEVICE EVALUATION: THE RETAIN CARTRIDGE INVOLVED WITH THIS COMPLAINT EVALUATED BY ANIMAS PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE CARTRIDGE HAS PASSED TESTING WITH NO FAULTS FOUND.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT CLAIMED SHE HAD DIFFICULTIES WITH CYCLING THE 4 ANIMAS CARTRIDGES FOR A MANUAL PRIME ON (B)(6) 2012. THAT SAME DAY, HER BLOOD GLUCOSE READING ELEVATED TO 510 MG/DL WHILE SHE HAD SYMPTOM OF NAUSEA AND TESTED POSITIVE FOR MODERN KETONES. THE PATIENT THEN WENT ON THE BACKUP PLAN. SHE TOOK NOVOLOG AND LANTUS INSULIN VIA SYRINGE IN THE EVENING. HER BLOOD GLUCOSE READING RESOLVED TO 140 MG/DL THE NEXT MORNING. THERE WAS NO PRODUCT MISUSE. THE PATIENT USED THE CARTRIDGE PER INSTRUCTION OF USE. THE CARTRIDGE CYCLE ISSUE WAS NOT RESOLVED WITH TRAINING. THE PATIENT PLANNED TO CONTINUE INSULIN PUMP THERAPY AS SOON AS SHE RECEIVED NEW LOT # OF CARTRIDGE. THE 4 ANIMAS CARTRIDGE IN QUESTION HAVE BEEN DISCARDED AND WILL NOT BE RETURN FOR INVESTIGATION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT EXPERIENCE ELEVATED BLOOD GLUCOSE OVER 510 MG/DL AND WAS TESTING POSITIVE FOR KETONES ON THE SAME DAY SHE HAD THE ALLEGED CARTRIDGE CYCLING ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM IR1200/1250/2020/OTP CART B201832

Patients

Seq Age Sex Outcome Treatment
1 36 YR Life Threatening| R