FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2761196 · Received September 26, 2012

Report

Report Number
3004209178-2012-89690
Event Type
Malfunction
Date Received
September 26, 2012
Date of Event
August 18, 2012
Report Date
September 19, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. UNABLE TO CONFIRM THE ERROR ALARM.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP HAD AN ERROR ALARM DURING THE MANUAL PRIME PROCESS. ADVISED THE CUSTOMER TO REVERT TO BACK UP PLAN. THE CUSTOMER'S BLOOD GLUCOSE READING WAS 167MG/DL. NO FURTHER INFORMATION WAS PROVIDED.