FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2761196
·
Received September 26, 2012
Report
- Report Number
- 3004209178-2012-89690
- Event Type
- Malfunction
- Date Received
- September 26, 2012
- Date of Event
- August 18, 2012
- Report Date
- September 19, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK. UNABLE TO CONFIRM THE ERROR ALARM.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP HAD AN ERROR ALARM DURING THE MANUAL PRIME PROCESS. ADVISED THE CUSTOMER TO REVERT TO BACK UP PLAN. THE CUSTOMER'S BLOOD GLUCOSE READING WAS 167MG/DL. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | OYC | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |