ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-11311
- Event Type
- Malfunction
- Date Received
- September 26, 2012
- Report Date
- September 16, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION (B)(4) 2012- DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE AUDIO BOLUS BUTTON WAS TORN. A DAMAGED BUTTON WILL ALLOW CONTAMINATION TO PERMEATE THE BUTTON CONTACT AND IMPACT THE BUTTON FUNCTION. DURING TESTING, THE AUDIO BOLUS BUTTON RESPONDED PROPERLY TO USER PRESSES. UNRELATED TO THE COMPLAINT, ALL THE KEYPAD BUTTONS WERE RESPONSIVE AND CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS.
ON (B)(6) 2012, THE PATIENT CONTACTED ANIMAS AND REPORTED THAT THE AUDIO BOLUS BUTTON WAS INTERMITTENTLY UNRESPONSIVE. REPORTEDLY, THE AUDIO BOLUS BUTTON WAS CRACKED AND TORN. THE PATIENT INDICATED THAT THE TACTILE CHANGES OCCURRED PRIOR TO THE DAMAGE. THE PATIENT DENIED EVIDENCE OF MOISTURE IN THE PUMP. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED DUE TO AN ALLEGED KEYPAD ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |