FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2761017 · Received September 26, 2012

Report

Report Number
1416980-2012-00787
Event Type
Malfunction
Date Received
September 26, 2012
Date of Event
June 27, 2012
Report Date
September 11, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Removal / Correction Number
1423500-01/08/10-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EC
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH EVENT HISTORY LOG REVIEW. THE CAUSE WAS DETERMINED TO BE AN UNEXPECTED HIGH ULTRA-FILTRATION (UF) FOR THIS THERAPY COMPARED TO OTHER THERAPIES. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2012 12:04:33. DURING NIGHT DRAIN CYCLE ONE, THE PATIENT'S ULTRAFILTRATION READING WAS 1893ML, INDICATING THE HOME PATIENT (HP) DRAINED 1893ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 190ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1