VITROS CHEMISTRY PRODUCTS PHBR SLIDES
Report
- Report Number
- 1319809-2012-00243
- Event Type
- Malfunction
- Date Received
- September 26, 2012
- Date of Event
- August 1, 2012
- Report Date
- September 26, 2012
- Product Code
- DLZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION DETERMINED THAT IMPRECISE VITROS PHBR QUALITY CONTROL RESULTS WERE OBTAINED ON A VITROS 5,1 FS CHEMISTRY SYSTEM. RESULTS OF PRECISION TESTING INDICATED THAT THE VITROS 5,1 FS CHEMISTRY SYSTEMS WERE OPERATING AS EXPECTED AT THE TIME OF THE EVENT. THE INVESTIGATION TO DATE HAS NOT IDENTIFIED A DEFINITIVE ROOT CAUSE, AND THE INVESTIGATION IS ONGOING. A REAGENT RELATED ISSUE OR AN INTERACTION BETWEEN THE VITROS PHBR SLIDES AND THE VITROS 5,1 FS CHEMISTRY SYSTEM HAVE NOT BEEN RULED OUT AS POTENTIAL CONTRIBUTING FACTORS. ACCEPTABLE QUALITY CONTROL RESULTS WERE OBTAINED AFTER USING AN ALTERNATE VITROS PHBR REAGENT LOT. THE ROOT CAUSE OF THIS EVENT IS CURRENTLY UNKNOWN. ADDITIONAL INVESTIGATION BY OCD REGARDING VITROS PHBR PERFORMANCE IS ONGOING.
A CUSTOMER OBTAINED IMPRECISE VITROS PHBR QUALITY CONTROL RESULTS ON A VITROS 5,1 FS CHEMISTRY SYSTEMS AS FOLLOWS: QC LOT K1912 (EXPECTED VALUE = 10.36 G/ML): 6.46 G/ML; QC LOT L1913 (EXPECTED VALUE = 27.40 G/ML): 17.54, 36.60 G/ML; QC LOT M1914 (EXPECTED VALUE = 57.89 G/ML): 43.77, 41.86, AND 42.13 G/ML. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF PATIENT SAMPLES WERE AFFECTED. NO PREVIOUSLY REPORTED VITROS PHBR PATIENT RESULTS WERE QUESTIONED WHILE QUALITY CONTROL RESULTS WERE OUTSIDE OF EXPECTED RANGES. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER ONE OF SIX MDRS FOR THIS EVENT. SIX 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SIX DEVICES WERE INVOLVED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITROS CHEMISTRY PRODUCTS PHBR SLIDES | IN-VITRO DIAGNOSTIC | DLZ | 2538-0060-8971 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |