ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2012-11272
- Event Type
- Malfunction
- Date Received
- September 26, 2012
- Report Date
- August 30, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP #1 SUBMITTED: (B)(4) 2012. DEVICE EVALUATION: A RETAINED CARTRIDGE SAMPLE FROM LOT # B201841 WAS EVALUATED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST, FILL TEST, AND FORCE TEST WAS PERFORMED WITH NO FAILURES OBSERVED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PATIENT CLAIMED HE MOISTURE DROPLETS INSIDE THE CARTRIDGE COMPARTMENT AND CARTRIDGE CHAMBER NEAR THE PLUNGER END. THE PATIENT ADMITTED HE USED COLD INSULIN AND WANTED TO KNOW IF THE MOISTURE DROPLETS ARE DUE TO CONDENSATION. THERE IS NO MISSING INSULIN. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO A POTENTIAL LEAKAGE IN THE CARTRIDGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN PUMP CARTRIDGE | LZG | ANIMAS CORPORATION | ANM IR1200/1250/2020/OTP CART | B201841 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |