FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2760622 · Received September 26, 2012

Report

Report Number
2531779-2012-11233
Event Type
Malfunction
Date Received
September 26, 2012
Report Date
August 29, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. CONTAMINATION AND DAMAGE WAS OBSERVED TO THE FORCE SENSOR ASSEMBLY. DURING EVALUATION THE PUMP WAS ABLE TO REWIND, LOAD AND PRIME SUCCESSFULLY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE

Description of Event or Problem · 1

THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 22 YR