PWRD ECH FLEX 60MM
Report
- Report Number
- 3005075853-2012-04352
- Event Type
- Malfunction
- Date Received
- September 26, 2012
- Date of Event
- July 31, 2012
- Report Date
- August 3, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K110385
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DAMAGED RELEASE BUTTON. THREE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ON WHAT TISSUE TYPE WAS THE DEVICE USED? AT WHAT LOCATION ON THE TISSUE? ON WHICH FIRING(S) DID THIS EVENT OCCUR (1ST, 2ND, 12TH, ETC.)? DURING WHICH STROKE DID THE EVENT OCCUR? WHAT COLOR CARTRIDGE WAS BEING USED? WHAT OTHER COLOR CARTRIDGES WERE USED BEFORE AND AFTER THIS EVENT? WAS BUTTRESSING MATERIAL UTILIZED? IF SO, WHICH PRODUCT? WAS THE INSTRUMENT FIRED ACROSS OR NEAR AN EXISTING STAPLE LINE OR CLIP? WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? WERE ANY OF THE FORCES HIGHER OR LOWER THAN EXPECTED (CLOSING, FIRING, OR OPENING)? AFTER USE, DID EACH OF THE TRIGGERS AND BUTTONS AUTOMATICALLY RETURN TO THEIR ORIGINAL (PRE-FIRED) POSITIONS, WITHOUT INTERVENTION? WAS THERE ANY DIFFICULTY REMOVING THE DEVICE FROM THE TISSUE? IS THERE ANY OTHER ADDITIONAL INFORMATION YOU WOULD LIKE TO SHARE ABOUT THIS DEVICE OR PROCEDURE? WHAT WERE THE PATIENT'S PRE-EXISTING CONDITIONS? DOES THE PATIENT HAVE ANY RELEVANT HISTORY OF SURGICAL TREATMENTS? IF SO, WHAT? HOW LONG HAS THE SURGEON BEEN USING THIS DEVICE FOR THIS PROCEDURE (IN YEARS)? WAS THERE A RECENT CONVERSION TO EES DEVICES IN THIS ACCOUNT OR WITH THIS SURGEON? THE ANALYSIS FOUND THAT ONE DEVICE WAS RECEIVED FOR ANALYSIS WITHOUT CARTRIDGE RELOAD. FURTHERMORE, THE DEVICE WAS FOUND TO HAVE THE CLAMPING MECHANISM DAMAGED. IN ORDER TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS, THE DEVICE WAS DISASSEMBLED. UPON DISASSEMBLING, THE CLAMP ARM WAS NOTED TO BE WORN OUT, IT APPEARS THAT THE DEVICE WAS FORCED OPEN WITH THE KNIFE WAS NOT ON THE HOME POSITION. THE INSTRUMENT WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT THE DEVICE WAS USED DURING A LAPAROSCOPIC SIGMOID RESECTION. ON THE 3RD FIRING THE DEVICE WOULD NOT FIRE. THE HANDLE WAS NOT CLOSING DOWN ALL THE WAY. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PWRD ECH FLEX 60MM | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | J4AZ21 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |