ARCHITECT I2000SR ANALYZER
Report
- Report Number
- 1628664-2012-00438
- Event Type
- Malfunction
- Date Received
- September 26, 2012
- Date of Event
- August 29, 2012
- Report Date
- September 6, 2012
- Manufacturer
- ABBOTT MANUFACTURING INC
- Product Code
- JJE
- PMA / PMN Number
- K983212
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, REVIEW OF THE INSTRUMENT LOG, A SEARCH FOR SIMILAR COMPLAINTS, AND A REVIEW OF LABELING. FROM THE TROUBLESHOOTING QUESTIONS, WE CONFIRMED IN THE INSTRUMENT LOG THAT IT WAS DECONTAMINATED WITH A BLEACH SOLUTION ON (B)(6) 2012, ONE DAY BEFORE THE EVENTS OCCURRED. WE INFORMED THE CUSTOMER THAT FALSE HIGH RESULTS COULD HAVE OCCURRED DUE TO THE FACT THAT THERE WAS SOME RESIDUE FROM THE DECONTAMINATION THAT COULD HAVE INTERFERED IN THE ASSAY. THE ISSUE WAS ADDRESSED BY REPLACING WASH ZONE MANIFOLD VALVES AND DECONTAMINATING THE SYSTEM. THE DECONTAMINATION WAS IDENTIFIED AS THE PRECURSOR TO THE ISSUE DESCRIBED IN THE CURRENT COMPLAINT. TRACKING AND TRENDING DID NOT IDENTIFY AN ADVERSE TREND. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. BASED ON THE AVAILABLE INFORMATION NO PRODUCT DEFICIENCY OF THE ARCHITECT I2000SR ANALYZER, LIST NUMBER 03M74 WAS IDENTIFIED.
(B)(4). AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER OBSERVED A FALSELY ELEVATED INSULIN RESULT WHILE USING THE ARCHITECT I2000SR ANALYZER. THE CUSTOMER PROVIDED THE FOLLOWING DATA: (B)(6) 2012: INITIAL 71.4UU/ML, THE RESULT WAS QUESTIONED BY THE PHYSICIAN AND RERUN ON (B)(6) 2012: 23.6 UU/ML. NO IMPACT TO PATIENT MANAGEMENT WAS DOCUMENTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT I2000SR ANALYZER | JJE | ABBOTT MANUFACTURING INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |