FDA Adverse Event Malfunction Summary report: N

M SERIES

MDR report key: 2760389 · Received September 18, 2012

Report

Report Number
1220908-2012-02499
Event Type
Malfunction
Date Received
September 18, 2012
Report Date
September 4, 2012
Manufacturer
ZOLL MEDICAL CORP
Product Code
MKJ
PMA / PMN Number
K990762
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORP HAS NOT RECEIVED THE PRODUCT FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING TRAINING BY FACILITY STAFF, THE DEVICE WAS UNABLE TO RESPOND TO INPUT CONTROLS. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORP MSERIES BI-PHASIC NA

Patients

Seq Age Sex Outcome Treatment
1 NA