FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 2760211
·
Received September 20, 2012
Report
- Report Number
- 1314492-2012-00294
- Event Type
- Malfunction
- Date Received
- September 20, 2012
- Date of Event
- August 22, 2012
- Report Date
- August 22, 2012
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). SIGMA EVALUATED THE DEVICE IN QUESTION. IT WAS CONFIRMED THAT THE AUDIO FUNCTION WAS NOT PRESENT. FURTHER EVAL IDENTIFIED THAT THE ABSENCE OF AUDIO WAS DUE TO A FAILED 1/2 WATT SPEAKER. ALL VISUAL FUNCTIONS PERFORMED AS EXPECTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PUMP DID NOT HAVE AUDIO FUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |