FDA Adverse Event Malfunction Summary report: N

GALILEO ECHO

MDR report key: 2760158 · Received September 26, 2012

Report

Report Number
1034569-2012-00177
Event Type
Malfunction
Date Received
September 26, 2012
Date of Event
August 29, 2012
Report Date
September 26, 2012
Manufacturer
IMMUCOR, INC.
Product Code
KSZ
PMA / PMN Number
BK070016
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

THE IMAGE RESULT FILES WERE REVIEWED BY AN IMMUCOR TECHNICAL SUPPORT SPECIALIST. CELL 1 OF THE SCREENING ASSAY IS POSITIVE FOR THE K ANTIGEN. THE SAMPLE RESULTED AS NEGATIVE FOR ALL CELLS. ALL CELLS VISUALLY APPEARED NEGATIVE AS REPORTED. CONTROL RESULTED AS EXPECTED. UPON REVIEW OF THE ANTIBODY IDENTIFICATION ASSAY RESULTS, A DIFFERENT SAMPLE (SAME PATIENT) WAS TESTED AND NEGATIVE RESULTS WERE OBTAINED FOR ALL CELLS. CELLS 8, 11 AND 14 (K POSITIVE CELLS) RESULTED AS NEGATIVE AND VISUALLY APPEARED NEGATIVE. THE CUSTOMER WAS ASKED TO REPEAT THE SAMPLE WITH A KNOWN POSITIVE SAMPLE. CELL 1, WHICH IS K POSITIVE, RESULTED AS NEGATIVE AND VISUALLY APPEARED NEGATIVE. CELL 2 RESULTED AS EQUIVOCAL. THE EXPECTED POSITIVE RESULTS WERE OBTAINED WITH THE KNOWN POSITIVE SAMPLE. AN ANTIBODY IDENTIFICATION PANEL WAS PERFORMED WITH A NEWLY DRAWN SAMPLE (SAME PATIENT) AND CELLS 1, 3, 6 AND 14 RESULTED AS POSITIVE (CELLS 8, 11 AND 14 ARE K POSITIVE). RESULTS VISUALLY APPEARED POSITIVE. A NEGATIVE RESULT WAS OBTAINED WITH CELL 8 WHICH VISUALLY APPEARED NEGATIVE. CELL 11 RESULTED AS EQUIVOCAL WITH THE RED CELL BUTTON HAVING A FUZZY APPEARANCE. CONTROLS RESULTED AS EXPECTED. THE UNEXPECTED REACTIVITY APPEARS TO BE SAMPLE RELATED.

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT UNEXPECTED NEGATIVE RESULTS WERE OBTAINED FOR ONE PATIENT ON THE CAPTURE-R READY ANTIBODY SCREENING AND IDENTIFICATION ASSAYS ON GALILEO ECHO INSTRUMENT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GALILEO ECHO AUTOMATED BLOOD BANK ANALYZER KSZ IMMUCOR, INC.

Patients

Seq Age Sex Outcome Treatment
1 62 YR