FDA Adverse Event
Malfunction
Summary report: N
TRANSTAR STRETCHER
MDR report key: 2760156
·
Received September 20, 2012
Report
- Report Number
- 1824206-2012-05704
- Event Type
- Malfunction
- Date Received
- September 20, 2012
- Date of Event
- August 31, 2012
- Report Date
- August 31, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FPO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND IF THE BRAKE IS SET FROM THE HEAD END, BOTH THE FOOT AND HEAD END PEDALS WILL SET INTO BRAKE. IT APPEARS THE BRAKE IS SET, HOWEVER, THE FOOT END BRAKE CASTERS WOULD NOT ENGAGE AND HOLD. THE TECHNICIAN ISOLATED THE ISSUE TO A BROKEN ROCKER ARM. HE USED A BRAKE/STEER UPGRADE KIT TO REPAIR THE STRETCHER.
Description of Event or Problem · 1
INFO RECEIVED INDICATES THE FOOT END CASTERS ARE NOT HOLDING WHEN THE BRAKE IS SET.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRANSTAR STRETCHER | WHEELED STRETCHER | FPO | HILL-ROM, INC. | 8000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |