FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 2760149
·
Received September 20, 2012
Report
- Report Number
- 1824206-2012-05712
- Event Type
- Malfunction
- Date Received
- September 20, 2012
- Date of Event
- August 30, 2012
- Report Date
- August 30, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THAT THE HI/LOW HEAD DOWN VALVE SEAT WAS STUCK IN PORT. HE REPLACED THE VALVE AND THE BED FUNCTIONED PROPERLY.
Description of Event or Problem · 1
INFO RECEIVED INDICATES THE HEAD END OF THE BED IS SLOWLY DRIFTING DOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |