FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2760149 · Received September 20, 2012

Report

Report Number
1824206-2012-05712
Event Type
Malfunction
Date Received
September 20, 2012
Date of Event
August 30, 2012
Report Date
August 30, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THAT THE HI/LOW HEAD DOWN VALVE SEAT WAS STUCK IN PORT. HE REPLACED THE VALVE AND THE BED FUNCTIONED PROPERLY.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE HEAD END OF THE BED IS SLOWLY DRIFTING DOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1