FDA Adverse Event
Death
Summary report: N
PHYSIO CONTROL LP-10
MDR report key: 275774
·
Received April 27, 2000
Report
- Report Number
- 275774
- Event Type
- Death
- Date Received
- April 27, 2000
- Date of Event
- April 18, 2000
- Report Date
- April 26, 2000
- Manufacturer
- PHYSIO CONTROL
- Product Code
- LDD
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- CA, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
AMBULANCE RESPONDED TO A PULSELESS 7 MONTH OLD INFANT. ADVANCED LIFE SUPPORT WAS INITIATED. WHILE TRANSPORTING THE INFANT TO THE HOSP, PT WAS PLACED ON A CARDIAC MONITOR. THE MONITOR DISPLAYED A "V-FIB" RHYTHM. MEDIC ATTEMPTED TWO (2) SHOCKS WITH NO CHANGE IN THE INFANT'S STATUS. INFANT DIED AT THE HOSP. THE MEDIC SUSPECTED A PROBLEM WITH THE MONITOR. THE MONITOR WAS EXAMINED BACK AT THE OPERATIONS BY THE MFR'S TECH, WHO CONCLUDED THAT THE PEDIATRIC PADDLES WERE NOT FUNCTIONING DUE TO A "BENT CONNECTOR." AS A RESULT, THE UNIT DISPLAYED A SHOCK, BUT NONE WERE EVER DELIVERED. BABY'S WEIGHT IS 7-10 LBS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PHYSIO CONTROL LP-10 | LIFE PAK EKG MONITOR | LDD | PHYSIO CONTROL | LP-10 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 MO | Death |