FDA Adverse Event Death Summary report: N

PHYSIO CONTROL LP-10

MDR report key: 275774 · Received April 27, 2000

Report

Report Number
275774
Event Type
Death
Date Received
April 27, 2000
Date of Event
April 18, 2000
Report Date
April 26, 2000
Manufacturer
PHYSIO CONTROL
Product Code
LDD
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
CA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

AMBULANCE RESPONDED TO A PULSELESS 7 MONTH OLD INFANT. ADVANCED LIFE SUPPORT WAS INITIATED. WHILE TRANSPORTING THE INFANT TO THE HOSP, PT WAS PLACED ON A CARDIAC MONITOR. THE MONITOR DISPLAYED A "V-FIB" RHYTHM. MEDIC ATTEMPTED TWO (2) SHOCKS WITH NO CHANGE IN THE INFANT'S STATUS. INFANT DIED AT THE HOSP. THE MEDIC SUSPECTED A PROBLEM WITH THE MONITOR. THE MONITOR WAS EXAMINED BACK AT THE OPERATIONS BY THE MFR'S TECH, WHO CONCLUDED THAT THE PEDIATRIC PADDLES WERE NOT FUNCTIONING DUE TO A "BENT CONNECTOR." AS A RESULT, THE UNIT DISPLAYED A SHOCK, BUT NONE WERE EVER DELIVERED. BABY'S WEIGHT IS 7-10 LBS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PHYSIO CONTROL LP-10 LIFE PAK EKG MONITOR LDD PHYSIO CONTROL LP-10 NA

Patients

Seq Age Sex Outcome Treatment
1 7 MO Death