INDEPENDENT CORNEAL VIEWING CHAMBER
Report
- Report Number
- 1119279-2012-00207
- Event Type
- Injury
- Date Received
- September 20, 2012
- Date of Event
- June 28, 2012
- Report Date
- August 20, 2012
- Manufacturer
- BAUSCH + LOMB
- Product Code
- LYX
- PMA / PMN Number
- K921729
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION RESULTS: THE LOT HISTORY RECORD AND THE STERILIZATION BATCH RECORD WERE REVIEWED FOR LOT NUMBER U7933 AND NO DISCREPANCIES OR DEVIATIONS THAT WOULD CONTRIBUTE TO THE REPORTED EVENT WERE NOTED. ALL END-PRODUCT RELEASE CRITERIA WERE MET. THE STERILIZATION CYCLE AND PARAMETERS MET THE ACCEPTANCE CRITERIA. OUR INVESTIGATION REVEALED THAT THE IVC-12 CORNEAL VIEWING CHAMBER WAS MANUFACTURED WITHIN SPECIFICATIONS AND THERE IS NO EVIDENCE OF ANY DEFECT IN THE STERILITY RECORDS. BASED ON THE CURRENT INFO, THE SPECIFIC ROOT CAUSE OF THE EVENT CANNOT BE DETERMINED. THIS EVENT HAS BEEN PREVIOUSLY REPORTED ON MANUFACTURER'S REPORT NUMBER 1119279-2012-00172 FOR OPTISOL-GS CORNEA STORAGE MEDIA.
IT WAS REPORTED THAT TWO PTS ((B)(6)) RECEIVED CORNEA TRANSPLANTS FROM THE SAME DONOR ((B)(6)). PT (B)(6) DEVELOPED ENDOPHTHALMITIS. ENTEROCOCCUS WAS IDENTIFIED ON THE DONOR CORNEA RIM CULTURE. PT WAS TREATED WITH VANCOMYCIN AND VITRECTOMY; HOWEVER, THE INFECTION DID NOT CLEAR AND ENUCLEATION WAS PERFORMED ON (B)(6) 2012. THIS REPORT IS FOR PT (B)(6) INVOLVING THE PT'S RIGHT EYE. PLEASE REFERENCE MDR #: 1119279-2012-00208 FOR PT (B)(6). ON (B)(4) 2012, ADDITIONAL INFO WAS RECEIVED INDICATING THAT THE CORNEAS HAD BEEN STORED IN IVC-12 CORNEAL VIEWING CHAMBERS. PER THE REPORTER, THE IVC-12 CORNEAL VIEWING CHAMBERS APPEARED TO BE SEALED PROPERLY, AND WERE DESCRIBED AS LOOKING GOOD. NO CULTURES WERE TAKEN PRIOR TO STORING THE CORNEAS IN OPTISOL GS CORNEA STORAGE MEDIA OR IVC-12 CORNEAL VIEWING CHAMBERS. THE REPORTER ALSO INDICATED THAT THERE WERE NO ROUTINE CULTURES PERFORMED UPON HARVESTING THE CORNEAS FROM THE DONOR AS CORNEAS ARE NON-STERILE TISSUE. THE RIM CULTURE OF CORNEA (B)(6) WAS DONE BY THE SURGEON AT THE TIME OF TRANSPLANT PER SURGEON'S ROUTINE PROCEDURE. THE PT'S PREOPERATIVE BCVA WAS COUNTING FINGERS. PT'S BCVA DECREASED TO HAND MOTION AS OF (B)(6) 2012.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INDEPENDENT CORNEAL VIEWING CHAMBER | INDEPENDENT VIEW CHAMBER | LYX | BAUSCH + LOMB | IVC-12 | U7933 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention| S | NON-STERILE HUMAN DONOR CORNEA| ANTI-REJECTION DROPS, AND| OPTISOL-GS CORNEA STORAGE MEDIA |