FDA Adverse Event
Other
Summary report: N
ANDREWS TABLE
MDR report key: 2757361
·
Received July 27, 2012
Report
- Report Number
- 2921578-2012-00014
- Event Type
- Other
- Date Received
- July 27, 2012
- Date of Event
- December 22, 2011
- Report Date
- July 24, 2012
- Manufacturer
- MIZUHO OSI
- Product Code
- FQO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TABLE AGE IS 20 YEARS, STATUS OF TABLE IS UNK. WILL CONTACT HOSPITAL FOR FURTHER DETAILS AND ADVISE ((B)(4) 2012).
Description of Event or Problem · 1
A PT WAS POSITIONED ON THE ANDREWS TABLE, THE TIBIAL SUPPORT LEVER WAS POSITIONED DOWN AND THE MOVEMENT LEVEL WAS PUSHED. ALL FUNCTIONS OF THE TABLE PROCEEDED TO MOVE, THE PT WAS REPOSITIONED AND THE LEVER WAS PUSHED AGAIN. ALL FUNCTIONS OF THE TABLE MOVED, EXCEPT IN A DIFFERENT DIRECTION. PT WAS REMOVED FROM TABLE AND REPOSITIONED ON A JACKSON TABLE AND SURGERY WAS CONTINUED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANDREWS TABLE | OPERATING ROOM TABLE | FQO | MIZUHO OSI | 5820 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |