FDA Adverse Event Malfunction Summary report: N

MANUAL CONTROLLED INFANT WARMER

MDR report key: 2757008 · Received September 24, 2012

Report

Report Number
9611451-2012-00668
Event Type
Malfunction
Date Received
September 24, 2012
Report Date
August 28, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
FMT
PMA / PMN Number
K971461
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: NO COMPLAINT DEVICE WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4). THEREFORE OUR EVALUATION IS BASED ON OUR KNOWLEDGE OF THE PRODUCT AND THE INFORMATION PROVIDED BY THE HOSPITAL. RESULTS: THE HOSPITAL REPORTED THAT THE HEATER HEAD IW990 INFANT WARMER UNIT WAS CRACKING. THE PHOTOGRAPHS PROVIDED BY THE CUSTOMER REVEALED CRACKING ON THE DOME PORTION OF THE HEATER HEAD CASE, ALONG WITH CRAZING AND FLAKING OF PLASTIC SHELL ON THIS AREA. PLASTIC CHIPPING WAS OBSERVED AT THE BOTTOM EDGE OF THE UPPER HEAD CASE. THE HOSPITAL REPORTED THAT THEY USED "CLINITEX" AS THE CLEANING AGENT FOR THE INFANT WARMER UNITS. A LOT CHECK REVEALED 6 OTHER COMPLAINTS OF THIS TYPE FROM THE SAME HOSPITAL FOR LOT NUMBER 070130. CONCLUSION: THE CRAZING AND CRACKING OF THE COMPLAINT WARMER HEADS ARE RELATED TO A KNOWN PROBLEM OF INCOMPATIBLE CLEANING SOLUTIONS REACTING WITH THE POLYCARBONATE PLASTIC AND ACRYLIC COMPONENTS OF THE INFANT WARMER. THE CLEANING AGENT USED BY THE HOSPITAL (CLINITEX), WHICH CONTAINS INGREDIENTS THAT ARE NOT RECOMMENDED BY FPH. WHEN THE HEATER HEAD BEGINS TO WARM UP DURING USE, A CHEMICAL REACTION WITH THE INCOMPATIBLE CLEANING SOLUTION CAN RESULT IN GRADUAL CRACKING AND CRAZING OF THE HEATER HEAD. CRAZING OR CRACKING USUALLY OCCURS AT THE AREAS WHERE INTERNAL STRESS HAS DEVELOPED DURING THE MOULDING OF THE HEAD CASE AND AGGRAVATION. THE INFANT WARMER SERVICE MANUAL FEATURES A DIAGRAM OF THE INFANT WARMER WHICH HIGHLIGHTS POLYCARBONATE AND ACRYLIC COMPONENTS AND STATES THE FOLLOWING: "CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYPOCHLORITES, PEROXIDES, GLUTARALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE." "CAUTION THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES". AS PART OF CONTINUOUS IMPROVEMENT, WE HAVE SINCE REVISED OUR CLEANING INSTRUCTIONS TO RECOMMEND SPECIFIC PROPRIETARY CLEANING WIPES. FPH REGIONAL OFFICE IN (B)(4) HAS SENT GUIDELINES ON HOW TO CLEAN THE PRODUCT AS PER THE IW900 TECHNICAL MANUAL. HEAD REPLACEMENT KIT WILL ALSO BE SENT TO THE HOSPITAL TO REPLACE THE DAMAGED HEATER HEAD.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT AN IW990 MANUAL CONTROLLED INFANT WARMER HAD CRACKING ON THE HEATER HEAD. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MANUAL CONTROLLED INFANT WARMER FMT FMT FISHER & PAYKEL HEALTHCARE LTD IW990

Patients

Seq Age Sex Outcome Treatment
1